TESEQ NSG 435 ESD SIMULATOR
Use our chat for personal support. Or contact us via +45 31 33 18 19 or salg@GOmeasure.dk
Compact and Cost Effective
- Programmable discharge voltage 200 V to 16.5 kV
- Stabilized charging voltage
- Pre-programmed IEC/EN 61000-4-2 test settings
- True air-discharge breakdown voltage
- Compact and lightweight design
Discover the possibilities
More information
Description
TESEQ NSG 435 ESD SIMULATOR
Ergonomic design and advanced functionality. The pistol-shaped NSG 435 simulator is designed to sit comfortably in the operator’s hand, with current operating conditions constantly displayed and clearly visible.
NSG 435’s microprocessor-based controller and multifunction keypad provide the operator with instant access to its comprehensive range of built-in functions. The LCD panel continuously displays the operating status and all user-selected test parameters.
NSG 435 has its own internal, battery operated high voltage generator. In addition to the pre-programmed, standard IEC pulses, the user can create custom tests using single or repetitive discharges with selectable rates and manual or automatic polarity switching. An optional mains supply is available for extended test operation or for use when the battery is being charged.
Parameter | Value |
---|---|
Basic set | Carrying case with: electrostatic discharge simulator, battery pack, battery charger 100 to 240 VAC, discharge network 150 pF/330 Ω, air- and contact-discharge tips, grounding cable, tripod adapter, user manual |
Pulsenetwork | Network 150 pF/330 Ω as per IEC/EN 61000-4-2 (included)Range of R/C networks for other standards: R = 0 Ω to 10 kΩ; C = 60 to 500 pF |
Discharge voltage | Air-discharge: 200 V to 16.5 kV (in 100 V steps)Contact-discharge: 200 V to 9 kV (in 100 V steps) |
Dischargetips | Ball and point as per IEC |
Charging voltage measurement | kV, accuracy better than ±5% (stabilized); measurement and display of true air flashover voltage |
Discharge detection | Indicated by an oval around the kV symbol, also acoustically in the 'Single' operating mode (air-discharge only) |
Holding time | >5 s (charging voltage ... ±5%) |
Polarity | Positive / negative / automatic change |
Operating modes | Single / repetitive =pulse counter: 0 to 9999, preselect counter 0 to 9999Repetition: 0.5/1/5/10/20 or 25 Hz (air); 0.5/1/5 or 10 Hz (contact);continuous operation |
Display | LCD panel showing: charging voltage, discharge voltage, polarity, air-/contact-discharge, counter/preselect counter content, soft-key functions, battery monitor |
Weight | NSG 435 with battery: 1.2 kg (2.6 lbs) approx. |
Ambient conditions | Operation: +5 to +40°C, 20 to 80 % r.h. (non-condensing), 68 to 106 kPa |
Documents
Options
Video
TESEQ NSG 435 ESD SIMULATOR
Ergonomic design and advanced functionality. The pistol-shaped NSG 435 simulator is designed to sit comfortably in the operator’s hand, with current operating conditions constantly displayed and clearly visible.
NSG 435’s microprocessor-based controller and multifunction keypad provide the operator with instant access to its comprehensive range of built-in functions. The LCD panel continuously displays the operating status and all user-selected test parameters.
NSG 435 has its own internal, battery operated high voltage generator. In addition to the pre-programmed, standard IEC pulses, the user can create custom tests using single or repetitive discharges with selectable rates and manual or automatic polarity switching. An optional mains supply is available for extended test operation or for use when the battery is being charged.
Parameter | Value |
---|---|
Basic set | Carrying case with: electrostatic discharge simulator, battery pack, battery charger 100 to 240 VAC, discharge network 150 pF/330 Ω, air- and contact-discharge tips, grounding cable, tripod adapter, user manual |
Pulsenetwork | Network 150 pF/330 Ω as per IEC/EN 61000-4-2 (included)Range of R/C networks for other standards: R = 0 Ω to 10 kΩ; C = 60 to 500 pF |
Discharge voltage | Air-discharge: 200 V to 16.5 kV (in 100 V steps)Contact-discharge: 200 V to 9 kV (in 100 V steps) |
Dischargetips | Ball and point as per IEC |
Charging voltage measurement | kV, accuracy better than ±5% (stabilized); measurement and display of true air flashover voltage |
Discharge detection | Indicated by an oval around the kV symbol, also acoustically in the 'Single' operating mode (air-discharge only) |
Holding time | >5 s (charging voltage ... ±5%) |
Polarity | Positive / negative / automatic change |
Operating modes | Single / repetitive =pulse counter: 0 to 9999, preselect counter 0 to 9999Repetition: 0.5/1/5/10/20 or 25 Hz (air); 0.5/1/5 or 10 Hz (contact);continuous operation |
Display | LCD panel showing: charging voltage, discharge voltage, polarity, air-/contact-discharge, counter/preselect counter content, soft-key functions, battery monitor |
Weight | NSG 435 with battery: 1.2 kg (2.6 lbs) approx. |
Ambient conditions | Operation: +5 to +40°C, 20 to 80 % r.h. (non-condensing), 68 to 106 kPa |