TESEQ NSG 4060 Series Low frequency immunity test system

Sale price0,00 DKK Price on quotation
Available

Use our chat for personal support. Or contact us via +45 31 33 18 19 or salg@GOmeasure.dk


  • Signal generator and integrated power amplifier 15 Hz to 150 kHz
  • Meets IEC / EN 61000-4-16 in combination with NSG 4060-1
  • Meets IEC / EN 61000-4-19 in combination with CDND M316-2 for voltage and CT 419-5 for current testing
  • Meets IEC / EN 60255-26 Annex A in combination with CN 60255-26
  • Prepared test configurations for IEC / EN 61000-4-16 and IEC / EN 61000-4-19
  • 5.7” color display
  • Multiple EUT monitoring options

Discover the possibilities

More information

TESEQ NSG 4060 Series Low frequency immunity test system



The NSG 4060 is an EMC immunity test system for the frequency range 15 Hz to 150 kHz. It consists of a sine wave generator, power amplifier, EUT monitoring interfaces and different coupling units depending on the application, i. e. NSG 4060-1 for testing IEC / EN 61000-4-16 including short duration disturbance tests for DC, AC from 162/3 Hz to 200 Hz and test levels up to 300 V. Voltage testing according IEC / EN 61000-4-19 requires the combination with CDND M316-2. It includes the differential mode coupling and decoupling and the required 10 Ω impedance of the disturbance source. Current testing according IEC / EN 61000-4-19, e.g. testing electricity meters, requires the combination with CT 419-5.
The powerful and easy to use firmware makes the NSG 4060 independent from an external PC and control software, however it can also be remote controlled for system operation. Part of NSG 4060’s delivery is an USB-to-serial / optical converter which offers potential free remote control.





Technical specifications NSG 4060 in combination with NSG 4060-1



Application:IEC 61000-4-16 edition 2.0, EN 61000-4-16:1998 / FprA3:2015
Frequency rangeShort duration test 1 s: DC, 162/3 Hz to 200 HzSweep and continuous test: 13 Hz to 150 kHz
Output voltageShort duration test 1 s: 0.1 Vrms to ≥300 Vrms (max. 2 A), 0.1 Vrms to 100 V DCSweep and continuous test: 0.1 Vrms to ≥30 Vrms
Phase synchronisation:
Harmonic distortion:Short duration test: <10%Sweep and continuous test: <1%
Rise and fall time of the DC output voltage at on / off switching:between 1 to 5 μs
Output impedance:50 Ω ±10%





Technical specifications NSG 4060 in combination with CDND M316-2



Application:IEC 61000-4-19 edition 1.0, EN 61000-4-19:2015
Frequency range:2 kHz to 150 kHz
Output voltage:0.1 Vrms to ≥20 Vrms
Output impedance:10 Ω ±30%
Harmonic distortion:<5%
Pulse modulationModulation frequency range: 3 Hz to 10 kHzFrequency resolution: 0.01 HzDuty cycle: 50%Rise / fall time (10% / 90%): < 10 μs





Technical specifications NSG 4060 in combination with CT 419-5



Application:IEC 61000-4-19 edition 1.0, EN 61000-4-19:2015TR 50579:2012
Frequency range:2 kHz to 150 kHz
Output current:2 kHz to 30 kHz 0.1 A to ≥4 A30 kHz to 150 kHz 0.1 A to ≥2 A
Harmonic distortion:<5%
Pulse modulation:Modulation frequency range: 3 Hz to 10 kHzFrequency resolution: 0.01 HzDuty cycle: 50%Rise / fall time (10% / 90%): < 10 μs
Output impedance: 1 Ω ±30%





General specifications NSG 4060



Display:640 x 480, 5.7” color
Output connector:4 mm safety banana
User port:D-Sub 15 pole with 4 TTL inputs and 4 TTL outputs+12 V / 100 mA, -12 V / 100 mA, +5 V / 100 mA power supply
Monitoring input analog:BNC socket, 0-24 V Ri=>15 kΩ, 6 mV resolution
Monitoring digital input:BNC socket, 0 to 24 V via optical coupler Ri=1.5 k Ω, switching threshold approx. 2 to 3 V
Monitoring optical input:optical fiber, HP versatile link HFBR0501 series 40 kBd
Sensor input:ODU socket, 0 to 5 V Ri≥4 kΩ, 6 mV resolution
External modulation input:BNC socket (prepared only)
RS232:D-Sub 9 pole, up to 115200 Bd
RS232 optical:Connector 2 x HFBRx523 socket for 1 mm fiber optic cable with length between 5 m and 30 m with 115200 Bd, for other distances 38400 Bd, max. 50 m
Mains input:110 to 230 VAC 50 / 60 Hz, autoranging
Size (W x H x D):45 cm (19“) x 15 cm (3HU) x 42.3 cm (with handle bar and foot)
Weight:approx. 15 kg
Cardboard box:60 cm x 55 cm x 37 cm, weight of cardboard box approx. 2.3 kg (empty)





General specifications NSG 4060-1 Extension Unit



Input / output connectors:4 mm safety bananas
Size (W x H x D):45 cm (19“) x 15 cm (3HU) x 37.5 cm
Weight:approx. 35 kg
Packing case:66 cm x 55 cm x 32 cm, approx. 10 kg (empty)





Technical specifications NSG 4060 in combination with NSG 4060-1 and NSG 4060-op1



Application:IEC 61000-4-16 edition 2.0, EN 61000-4-16:1998 / FprA3:2015
Frequency range
Short duration test 1 s:DC
Output voltage
Short duration test 1 s:1 V to 330 V DC
Continuous test:1 V to ≥30 V DC
Resolution:0.01 V
Rise and fall time of the DC output voltage at on / off switching:between 1 to 5 µs
Output impedance:50 Ω ±10%





General specifications NSG 4060-op1



Mains input:200 to 240 VAC 47 / 63 Hz
Size (W x H x D):43 cm x 9 cm x 43 cm
Weight:approx. 13 kg