Hioki 8967 Memory HiCorder Temp Unit
Use our chat for personal support. Or contact us via +45 31 33 18 19 or salg@GOmeasure.dk
TEMP UNIT8967
Measure temperature waveform by 16 bit resolution, isolated input
Discover the possibilities
More information
Description
Key Features
• 16-bit A/D high resolution measurement
• Isolated input, up to 300 V AC/DC from the earth
Basic specifications (Accuracy guaranteed for 1 year, Post-adjustment accuracy guaranteed for 1 year)
Measurement functions | Number of channels: 2, for temperature measurement with thermocouple (voltage measurement not available) |
---|---|
Input connectors | Thermocouple input: plug-in connector,Recommended wire diameter: single-wire, 0.14 to 1.5 mm2, braided wire 0.14 to 1.0 mm2 (conductor wire diameter min. 0.18 mm), AWG 26 to 16Input impedance: min. 5 MΩ (with line fault detection ON/OFF),Max. rated voltage to earth: 300 V AC, DC(with input isolated from the unit, the maximum voltage that can be applied between input channel and chassis and between input channels without damage) |
Temperature measurement range Note: Upper and lower limit values depend on the thermocouple | 10 °C/div (-100 °C to 200 °C), 50 °C/div (-200 °C to 1000 °C), 100 °C/div (-200 °C to 2000 °C), 3 ranges, full scale: 20 div,Measurement resolution: 1/1000 of measurement range (using 16-bit A/D conversion and when installed in the 8847) |
Thermocouple range (JIS C 1602-1995) (ASTM E-988-96) | K: -200 to 1350 °C, J: -200 to 1100 °C, E: -200 to 800 °C, T: -200 to 400 °C, N: -200 to 1300 °C, R: 0 to 1700 °C, S: 0 to 1700 °C,B: 400 to 1800 °C, W (WRe5-26): 0 to 2000 °C,Reference junction compensation: internal/ external (switchable),Line fault detection ON/OFF possible |
Data refresh rate | 3 methods, Fast: 1.2 ms (digital filter OFF), Normal: 100 ms (digital filter 50/60 Hz), Slow: 500 ms (digital filter 10Hz) |
Measurement accuracy | Thermocouple K, J, E, T, N: ±0.1 % of full scale ±1 °C (±0.1 % of full scale ±2 °C at -200 °C to 0 °C),Thermocouple R, S, W: ±0.1 % of full scale ±3.5 °C (at 0 °C to 400 °C or less), ±0.1 % of full scale ±3 °C (at 400 °C or more)Thermocouple B: ±0.1 % of full scale ±3 °C (at 400 °C or more),Reference junction compensation accuracy: ±1.5 °C (added to measurement accuracy with internal reference junction compensation) |
Dimensions and mass | 106 mm (4.17 in) W × 19.8 mm (0.78 in) H × 204.5 mm (8.05 in) D, 240 g (8.5 oz) |
Accessories | Ferrite clamp × 2 |
Documents
Options
Video
Key Features
• 16-bit A/D high resolution measurement
• Isolated input, up to 300 V AC/DC from the earth
Basic specifications (Accuracy guaranteed for 1 year, Post-adjustment accuracy guaranteed for 1 year)
Measurement functions | Number of channels: 2, for temperature measurement with thermocouple (voltage measurement not available) |
---|---|
Input connectors | Thermocouple input: plug-in connector,Recommended wire diameter: single-wire, 0.14 to 1.5 mm2, braided wire 0.14 to 1.0 mm2 (conductor wire diameter min. 0.18 mm), AWG 26 to 16Input impedance: min. 5 MΩ (with line fault detection ON/OFF),Max. rated voltage to earth: 300 V AC, DC(with input isolated from the unit, the maximum voltage that can be applied between input channel and chassis and between input channels without damage) |
Temperature measurement range Note: Upper and lower limit values depend on the thermocouple | 10 °C/div (-100 °C to 200 °C), 50 °C/div (-200 °C to 1000 °C), 100 °C/div (-200 °C to 2000 °C), 3 ranges, full scale: 20 div,Measurement resolution: 1/1000 of measurement range (using 16-bit A/D conversion and when installed in the 8847) |
Thermocouple range (JIS C 1602-1995) (ASTM E-988-96) | K: -200 to 1350 °C, J: -200 to 1100 °C, E: -200 to 800 °C, T: -200 to 400 °C, N: -200 to 1300 °C, R: 0 to 1700 °C, S: 0 to 1700 °C,B: 400 to 1800 °C, W (WRe5-26): 0 to 2000 °C,Reference junction compensation: internal/ external (switchable),Line fault detection ON/OFF possible |
Data refresh rate | 3 methods, Fast: 1.2 ms (digital filter OFF), Normal: 100 ms (digital filter 50/60 Hz), Slow: 500 ms (digital filter 10Hz) |
Measurement accuracy | Thermocouple K, J, E, T, N: ±0.1 % of full scale ±1 °C (±0.1 % of full scale ±2 °C at -200 °C to 0 °C),Thermocouple R, S, W: ±0.1 % of full scale ±3.5 °C (at 0 °C to 400 °C or less), ±0.1 % of full scale ±3 °C (at 400 °C or more)Thermocouple B: ±0.1 % of full scale ±3 °C (at 400 °C or more),Reference junction compensation accuracy: ±1.5 °C (added to measurement accuracy with internal reference junction compensation) |
Dimensions and mass | 106 mm (4.17 in) W × 19.8 mm (0.78 in) H × 204.5 mm (8.05 in) D, 240 g (8.5 oz) |
Accessories | Ferrite clamp × 2 |