EM TEST CNV 504N5 Coupling/Decoupling Networks
Use our chat for personal support. Or contact us via +45 31 33 18 19 or salg@GOmeasure.dk
- For slow damped oscillatory waves as per EN/IEC 61000-4-18
- Coupling capacitor 0.5uF
- Decoupling inductor 1.5 mH for 50 V AC/DC
- Nominal line current 4A AC/DC
- Source impedance 200 Ohm (100 + 100 Ohm)
- max. 3kV peak for 100kHz and 1MHz damped oscillatory waves
Discover the possibilities
More information
Description
EM TEST CNV 504N5 Coupling/Decoupling Networks
The EM TEST CNV 504N5 Series is a coupling/decoupling networks are used to apply slow damped oscillatory waves onto signal/data lines in accordance to EN/IEC 61000-4-18. A capacitance of 0.5uF is specified to couple the pulses on to the lines. Each line has a decoupling inductor of 1.5mH or 3.0 mH to protect the auxiliary equipment from being subjected to the test pulse. The disturbance signal used for this test is a damped oscillatory pulse with a frequency of 100kHz or 1MHz, being generated by the EM TEST OCS 500N6 or OCS 500N6F. The CNV 504N5 series supports voltages of up to 250V AC/DC and a current of up to 4A AC/DC.Features
- Dedicated CDN for the coupling of slow damped oscillatory waves as per EN/IEC 61000-4-18
- Coupling capacitor 0.5uF
- Decoupling inductor 1.5 mH for 50 V AC/DC
- Decoupling inductor 3.0 mH for 250V AC/DC
- Nominal line current 4A AC/DC
For more information: Ametek
Documents
EM TEST CNV 504N5 Coupling/Decoupling Networks
EM TEST CNV 504N5 DatasheetOptions
Ametek CNV 504N5 Coupling/Decoupling Networks
AVAILABLE CNV 504N5 MODELS
CNV 504N5.1 : 50V AC/DC per line CNV 504N5.3 : 250V AC/DC per lineVideo
EM TEST CNV 504N5 Coupling/Decoupling Networks
The EM TEST CNV 504N5 Series is a coupling/decoupling networks are used to apply slow damped oscillatory waves onto signal/data lines in accordance to EN/IEC 61000-4-18. A capacitance of 0.5uF is specified to couple the pulses on to the lines. Each line has a decoupling inductor of 1.5mH or 3.0 mH to protect the auxiliary equipment from being subjected to the test pulse. The disturbance signal used for this test is a damped oscillatory pulse with a frequency of 100kHz or 1MHz, being generated by the EM TEST OCS 500N6 or OCS 500N6F. The CNV 504N5 series supports voltages of up to 250V AC/DC and a current of up to 4A AC/DC.Features
- Dedicated CDN for the coupling of slow damped oscillatory waves as per EN/IEC 61000-4-18
- Coupling capacitor 0.5uF
- Decoupling inductor 1.5 mH for 50 V AC/DC
- Decoupling inductor 3.0 mH for 250V AC/DC
- Nominal line current 4A AC/DC
For more information: Ametek